The Fisher Cups have all the advantages associated with a precision laboratory cup. The are Standard and Dip Cups versions available and they have removable orifices that are interchangeable, with a “snap in, snap out” type, permitting coverage of the complete viscosity range with one cup. The can also be replaced if damaged rather than requiring replacement of the entire cup.
Features
The Fisher series of viscosity cups as originally designed and produced, provided some advantages over other existing dip viscosity cups. They accommodated a larger sample of the test material. The orifice was considerably longer than its diameter resulting in cup efflux time more closely proportional to viscosity. However, there were deficiencies in the design. The cup was difficult to clean. It was costly to produce and to calibrate. A damaged or enlarged orifice resulted in necessity for replacement of the entire cup.
The Gardco/Fisher viscosity cups retain the original internal cup dimensions and the same orifice length and diameter but with the following improvements: The weight of the new cup has been reduced to 44% of the weight of the original cup for easier temperature control when cup and material are not at the same temperature. The orifices are of the readily replaceable type permitting coverage of the complete viscosity range with one cup and interchangeable orifices. The orifice may be readily replaced if damaged rather than requiring replacement of the entire cup. The design of the lower portion of the cup eliminates hard to clean areas. The design directs the flow of all test material from the outside of the cup away from the efflux stream improving cup accuracy. Rod supports are designed and secured to the cup in a manner to prevent drainage to the inside of the cup and to provide easier cleaning. In addition to physical improvements, the cups are furnished with conversion formulas and a table which relates the nearest tenth of a second of total efflux time to viscosity in centistokes. This table is particularly useful in determining efflux time in seconds when viscosity in centistokes is known.
The Gardco/Fisher viscosity cups are calibrated at time of manufacture to within 2% of target efflux time. This calibration is with standard oils traceable to the National Institute of Standards and Technology. At an additional cost any cup/orifice combination is available with a CERTIFICATE OF CALIBRATION. The procedures and conditions of calibration on which the certificate is based qualify under ANSI/NCSL Z540 or ISO/IEC 17025, ISO 9001, as applicable, & conforms to ISO 9000 when ordered with certification.
The Gardco/Fisher cups are produced to very close mechanical tolerance in elaborate jigs and fixtures. Such equipment not only insures that each cup is correctly produced but also that all cups are identical. All parts of the cup are of stainless steel except for the orifice and the name plate.
Technical Attributes
Orifice No. | Seconds Range | Centistoke Range | Midrange Sensitivity¹ | Calibration Oil No./Centistokes² |
1 | 22 - 60 | 11 - 48 | 0.9 | G 20 / 35 |
2 | 14 - 60 | 19 - 136 | 2.5 | G 35 / 68 |
3 | 10 - 60 | 35 - 320 | 10.3 | G 60 / 117 |
4 | 10 - 60 | 167 - 1125 | 19.0 | G 200 / 468 |
¹ Centistoke Values are Nominal - Actual Values Printed on Bottle Labels. ² Stated as Centistokes per Second of Efflux Time. |
More Information
All prices are F.O.B. Pompano Beach and are subject to change without notice.
Fisher Dip Cup Construction
The body of the cup is machined from solid bar stock stainless steel. Note the conical opening for the orifice which permits gentle “tapping in” or “tapping out” of the orifice. Also note that the cup shape insures that materials draining from the exterior of the cup flow away from the orifice. Stainless steel rods supporting the cup are formed so that they are held a distance away from the cup. The pad formed at the end of the support rod is welded to the cup below the cup rim. This design prevents any test material from the rod flowing to the inside of the cup. The increased separation of the two support rods also makes it easier to clean the cup. The orifices for the cup are machined from brass with the exact conical exterior to match the opening in the base of the cup. Positive orifice identification is assisted by “steps” at the base of the orifice.
Fisher Standard Cup Construction
The body of the cup is machined from solid bar stock stainless steel. Note the conical opening for the orifice which permits gentle “tapping in” or “tapping out” of the orifice. The orifices for the cup are machined from brass with the exact conical exterior to match the opening in the base of the cup. Positive orifice identification is assisted by “steps” at the base of the orifice.
Cup Temperature
Most materials change in viscosity as a function of temperature. Those normally measured by viscosity cups change in the range of 3% to 8% per degree Celsius change in temperature. Usually, the higher the viscosity the greater the change. For acceptable accuracy, it is necessary to measure temperature at the same time that viscosity cup readings are taken. When many determinations are to be made on similar products in the same viscosity range, it may be helpful to produce a graph for converting measured temperature and viscosity cup efflux time in seconds to seconds at a specified temperature, normally 25° Celsius.
There are three variables to consider: viscosity, efflux time seconds and temperature. All three can be shown on a graph with a family of curves as shown in the following example. Viscosity level is indicated by the diagonal lines, increasing from the lower left to the upper right. Such a graph can be prepared for a given material by taking readings with the Gardco/Fisher cup over a limited temperature range as shown in the example. Within this limited range the plots of the obtained data will normally result in a straight line such as the heavy diagonal line. Draw parallel lines to the line of experimental data, as shown, representing different viscosity levels. Enter on the graph the material represented and the Gardco/Fisher cup number designation.
Procedure
It is helpful and results will be more consistent if all viscosity determinations are made in a room that is free from drafts or quick temperature change and is within ±3° C ( ±5° F ) of the viscosity measuring temperature of 25° C.
Fisher Dip Cup Instructions
Fisher Standard Cup Instructions
Delivery Content
Instruction Manual and Grid Set provided with each cup. |
Accessories
Other Variations