+1 (954) 946-9454

EZ™ Zahn Viscosity Cups

Article No. 13789

Zahn EZ Viscosity Cup #3

  • EZ Zahn Viscosity Dip Cup #3
  • 64-596 Centistoke Range

Viscosity Calculator App

Download Excel Viscosity Chart HERE

Select your model

$173.00 excl. taxes & shipping cost

Features

  • The EZ™ Viscosity Cups are produced to very close mechanical tolerance in elaborate jigs and fixtures, ensuring that each cup is correctly produced and are identical
  • Made from Stainless Steel
  • In compliance to ASTM D4212
  • Calibration and Certification procedures qualify under ISO/IEC 17025 as applicable

Technical Attributes

  • Data
    Cup NumberSeconds RangeCentistoke RangeMidrange Sensitivity ¹Calibration Oil Number ²
    140 to 6010 to 361.3G-10/19
    220 to 6019 to 1563.3G-60/117
    312 to 6064 to 59610.5G-200/458
    410 to 6079 to 78413.9G-200/458
    510 to 60161 to 140124.2G-350/878
    ¹ Stated as centistokes per second of efflux time.
    ² Centistoke values are nominal - actual values printed on labels

Standards

  • ASTM D816 
  • ASTM D1084 
  • ASTM D4212

Procedure

  1. Select the proper number cup to be used from the Specification Table, which is dependent on the expected viscosity range of the material to be measured.
  2. Insure that the cup is clean and that there is no residual dried material in or around the orifice.
  3. Adjust the temperature, if necessary, of the test material.
  4. Completely immerse the cup into the material to be measured in a location free from bubbles or foam, holding the cup vertically by means of the stainless steel split key ring.
  5. Measure and record the temperature of the material that is encompassed by the cup.
  6. Hold cup vertically by inserting index finger into handle ring. In a quick, steady motion, lift the cup out of the sample material, starting the timer when the top edge of the cup breaks the surface. During the flow time, hold the cup no more than 6” above the level of the sample material.
  7. Stop the timer when the first definite break in the stream at the base of the cup is observed.
  8. Record the number of seconds of efflux time, temperature and the cup number. (Example: No. 2 EZ™ Dip Cup, 48.1 seconds at 25.1°C.) As an option to the preceding step, refer to the Viscosity Calculator App or conversion table for the cup to determine the centistoke viscosity for the measured efflux time in seconds and record this value and the measured temperature.
  9. Promptly clean the cup unless it will be used immediately for a rerun of the same material. (Use a length of nylon fishing line to clean the orifice.)

Knowledge

Want to learn the perfect process for using a viscosity cup? The article link below provides information on how to choose the right viscosity cup, how to measure efflux time, how to remove calibration oils and much more. - The Proper Process for Measuring with a Viscosity Cup

Join our Email List. subscribe