Features
Technical Attributes
Type of Detector | SJ-220 Standard Drive Unit | SJ-310 Standard Drive Unit | |||
Part No.: | 15507 | 15508 | 13130 | ||
Measurement: | 0.75mN | 4mN | 0.75mN | ||
Traverse Length: | - | .7" (17.5 mm) | |||
Measuring: (Detector) | Range: | 360 µm (-200 µm to +160 µm) / 14400 µin (-7900 µin to +6300 µin) | |||
Range/Resolution: | 14400 µin / .8 µin (360 µm / 0.02 µm) 4000 µin / .2 µin (100 µm / 0.006 µm) 1000 µn / .08 µin (25 µm / 0.002 µm) | ||||
Tracing Speed: | Measuring 0.01, 0.02, 0.03 in/s (0.25mm/s, 0.5mm/s, 0.75mm/s) Returning 1 mm/s | ||||
Measure Force/Stylus Tip: | 0.75 mN type: 0.75mN/2μuR 60°, 4mN type: 4mN / 5μmR 90° | ||||
Skid Force: | Less than 400mN | ||||
Assessed Profiles: | Primary profile / Roughness profile / DF profile / Roughness profile-Motif | ||||
Evaluation Parameters: | Ra, Rc, Ry, Rz, Rq, Rt, Rmax*1, Rp, Rv, R3z, Rsk, Rku, RPc, Rsm, Rz1max*2, S, HSC, RzJIS*3, Rppi, RΔa, RΔq, Rir, Rmr, Rmr(c), Roc, Rk, Rpk, Rvk, Mr1, Mr2, A1, A2, Vo, Rpm, tp, Htp*4, R, Rx, AR, Possible Customize | ||||
Analysis Graphs: | Bearing area curve / Amplitude distribution curve | ||||
Filters: | Gaussian, 2CR75, PC75 | ||||
Cut Off Length: | λc | 0.003, 0.01.0.03, 0.1” (0.08, 0.25, 0.8, 2.5mm) | |||
λs*5 | 100, 300μin (2.5, 8μm) | ||||
Sampling Length: | 0.003, 0.01, 0.03, 1” (0.08, 0.25, 0.8, 2.5mm) | ||||
Number of Sampling Lengths: (xn) | x1, x2, x3, x4, x5, x6, x7, x8, x9, x10, Arbitrary .0118 ~ .22"(.0001" Interval) [(0.3 ~ 5.6mm: 0.01mm Interval)] | ||||
LCD Dimensions: | 2.8" TFT colo LCD touch panel with backlight (320x240 dot) | 4.64" x 3.47" (117.8 x 88.2 mm) | |||
Display Languages: | Japanese, English, German, French, Italian, Spanish, Portuguese, Korean Traditional Chinese, Simplified Chinese, Czech, Polish, Hungarian, Turkish, Swedish, Dutch, Russian, Slovenian, Romanian, Bulgarian, Finnish, Thai, Vietnamese, Indonesian and Malay | Japanese, English, German, French, Italian, Spanish, Portuguese, Korean Traditional Chinese, Simplified Chinese, Czech, Polish, Hungarian, Turkish, Swedish, Dutch | |||
Measurement Result Display: | Vertical display: 1 parameter / 3 parameter / trace to measurements Horizontal display: 1 parameter / 4 parameter / trace to measurements (Horizontal display is invertable) | 1-parameter display: one parameter measurement result 4-parameter display: four parameter measurement results Profile display: one parameter measurement result and the measured profile Trace display: The ten latest measurement results using the same parameter | |||
Printing function*6: (Dedicated printer is required separately) | Measurement conditions / Calculation results / Calculation results for each sampling length / Assessed profile / Bearing area curve / Amplitude distribution curve / Environment setting information | ||||
External I/O: | USB I/F, Digimatic Output, Printer Output, RS-232C I/F, Foot SW 1/F | ||||
Functions: | Customization | Desired parameters can be selected for calculation and display | |||
GO/NG judgment*7 | By max value / 16% / Standard deviation | ||||
Storage of Measurement Condition | Save the condition at power OFF | ||||
Storage | Internal memory: Measurement condition (10 sets), Measured profile (1 set), Memory card (Option): 500 measurement conditions, 10000 measured profiles, 500 display images, Text file (Measurement conditions / Measured profile / Assessed profile / Bearing area curve / Amplitude distribution curve) | ||||
Calibration: | Auto-calibration with the entry of numerical value / Average calibration with multiple measurement (Max 5 time) is available | Auto-calibration with the entry of numerical value / Average calibration with multiple measurement (Max 12 time) is available | |||
Power-Saving Function: | Auto-sleep off function (10-600sec) *8 | Auto-sleep function (30-600 sec) | |||
Power Supply: | Two-way power supply: baqttery (rechargeable Ni-MH battery) and AC adapter *Charging time: about 4 hours (may vary due to ambient temperature) *Endurace: about 1000 measurements (differs slightly due to use conditions / environment) | Two-way power supply: battery (rechargeable Ni-MH battery) and AC adapter Note 1: Charging time: about 4 hours (may vary due to ambient temperature) Note 2: Endurance: about 1500 measurements (differs slightly due to use conditions /environment) | |||
Size: W x D x H | Display Unit: | 164.7 x 67.1 x 51.9 mm | 10.8" x 4.3" x 7.8" (275 x 109 x 198 mm) | ||
Drive Unit: | 115 x 23 x 26 mm | 4.5"x 0.9" x 8.9" (115 x 23 x 26.7 mm) | |||
Mass: | About 1.1 lbs. (500 g) (Display unit + Drive unit + Standard detector) | About 3.9 lbs. (1.8 kg) (Display unit + Drive unit + Standard detector | |||
Standard Detector | |||||
Measuring Force: | 0.75 mN | 4 mN | |||
Stylus Tip: | Radius 2 µm | Radius 5 µm | |||
Tip Angle: | 60° | 90° | |||
*1 Calculation is available only when selecting the VDA, ANSI, or JIS '82 standard. *2 Calculation is available only when selecting the ISO '97 standard. *3 Calculation is available only when selecting the JIS '01 standard. *4 Calculation is available only when selecting the ANSI standard. *5 Not available when selecting the JIS '82 standard. *6 Order the SJ-210 printer (178-421, optional accessory) separately. See page 19 for details about the SJ-210 printer. *7 Standard deviation only can be selected in ANSI. 16 % rule cannot be selected in VDA. *8 Auto-sleep function is invalid when AC adaptor is used. |
Standards
JIS (JIS-B0601-2001, JIS-B0601-1994, JIS B0601-1982), VDA, ISO- 1997, and ANSI
Delivery Content
Downloads