This FORM can help you design the Test Chamber you require
Features
Available in sizes ranging from 9.3 cu. ft. to 2,508 cu. ft., Singleton SCCH accelerated corrosion test chambers can be used for Salt Fog, CASS, Humidity, Water Fog and many other test methods.They are designed to provide the accuracy and flexibility necessary to meet today's demanding test methods. All Singleton testing chambers are easy to use and operate.
Technical Attributes
Model # | SCCH20 | SCCH21 | SCCH22 |
Capacity | 9.3 ft3 (264 liters) | 18 ft3 (510 liters) | 30 ft3 (850 liters) |
Inside Dimensions Exposure Zone Depth | 30" x 18" x 30" (76 x 45 x76 cm) | 36" x 24" x 36" (92 x 61 x 92 cm) | 48" x 30" x 36" (122 x 76 x 92 cm) |
Outside Dimensions (cover closed) | 50" x 28" x 48" (62" open) 127 x 72 x122 cm (158 open) | 56" x 34" x 56" (74" open) 142 x 86 x142 cm (188 open) | 65" x 36" x 56" (78" open) 165 x 92 x 145 cm (198 open) |
Electical Requirements: | 120V, 240V | 120V, 240V 480V | 240V, 240V-3ph or 480-3ph |
Temperature Range: | Ambient to 120°F (±3°) / or 50°C (±2°) | ||
Humidifying Tower: | Ambient to 140°F (±3°) / or 60°C (±2°) | ||
Humidity Range: | Ambient to 100% | ||
Fog Rate: | 1-3 ml/hr | ||
Supply Air Pressure: | 40-120 psi maximum | ||
Supply Air Volume: | 40-50 scfh @ 12-18 psi | ||
Drain: | Access to floor drain | ||
Exhaust: | Vent to the outside | ||
Warranty: | One year limited warranty |
Utility Requirements | |||
City Water: | All | Should be made available with the use of a shutoff valve and hose arrangement in the test area near the unit. Up to 100 gallons needed at initial installation / start up | |
Deionized Water (ASTM D1193 Type IV): | All | Used for filling reservoirs and for electrolyte solution makeup (up to 6 gallons per day) | |
Exhaust: | SCCH20-22 | 1” npt | |
Drain: | All | 1/2” npt | Corrosion resistant piping (ie) PVC |
More Information
Delivery Content
Accessories