+1 (954) 946-9454

RotoThinner

Article No. 13761

RotoThinner

The flexibility of the TQC Sheen Digital Rotothinner™, DV2700, allows the user the ultimate control during the viscosity reduction process whilst also providing very useful stability analysis. Spindle not included, see accessories. 

$3,995.00 excl. taxes & shipping cost

Features

  • Offers easy, continuous monitoring of the material viscosity during the addition of solvents and thinners to create supply or RFU viscosity.
  • Allows definitive viscosity analysis at times preselected by the user, providing excellent data relating to shear performance
  • Ultimate control during the viscosity reduction process whilst also providing very useful stability analysis.
  • Microprocessor controlled which ensures a higher accuracy and repeatability. Enhanced by the automatic multi-point calibration, for which no dongle is required.
  • Perfect for use in both production and laboratory environments.
  • The manual reading starts when the handle is at its lowest point.
  • The continuous reading gives continuous measurement results.
  • The max hold function shows the maximum reading.
  • The time-to-reading function offers the opportunity to automatically take readings with a set time interval.
  • An automatic display hold function is available to freeze the reading so it can be noted more easily.
  • Three measuring ranges achievable in one instrument
    • The plate spindle is used to determine viscosity of fluids of 0-22 poise.
    • The large ball spindle is used for measurements between 0-75 poise.
    • The small ball spindle is used for measurements between 0-350 poise.
  •  Conforming to ISO 2884 and BS 3900 A7 a range of verification oil kits are also available.

Technical Attributes

  • Data
    Mains:100-240VAC / 50-60HZ
    Operating Temperature:+15°C - +35°C / +59°F - + 95°F (Other temperature after proper calibration)
    Net Weight:8500 G / 18.7 LBS
    Speed:562 RPM
    Spindle:Supplied separately

    Range:
    Spindle 1: 0 - 22P 0 - 2200cP
    Spindle 2: 0.1 - 75P 10 - 7500cP
    Spindle 3: 1 - 350P 100 - 35000cP
    Resolution:0.1P
    1cP
    Accuracy:± 1% of full scale
    Repeatability:± 1% of full scale
    Sample Container:RL seal required for spindle 1
    Dimensions:200 x 360 x 550mm / 7.9 x 14.2 x 21.7inch (w x d x h)

Procedure

  1. A tin container is filled with sample fluid and positioned on a magnetic ring mounted on top of the base.
  2. The spindle is submersed in the sample fluid by lowering the handle.
  3.  When the handle has reached the lowest point, the spindle starts rotating with a set constant speed.
  4.  The display shows the actual measured viscosity in poise.
  5. The reading of free-flowing fluids can be viewed quickly. (More structured materials need more time because the spindle applies shear that influences the reading.)
  6. After taking the reading(s) the handle can be lifted to a level where the spindle is just below the rim of the sample container. This level offers the operator the opportunity to spin off the sample material.
  7. When the handle is raised to its maximum extent the spindle automatically switches off.
  8. The quick release chuck allows for quick spindle exchange and eases cleaning the instrument and spindles.