Features

An optical-mechanical gauge such as the Paint Inspection Gauge, if set up correctly, will not be subject to appreciable calibration change except by gradual wear of cutters and guide studs. On the other hand, “surface proximity gages”, magnetic, inductance, or eddy current types generally require calibration before any series of measurements; and frequently within a series of measurements - especially if the surfaces are variable in type and structural thickness.

  • All plastic film standards supplied as individual chips
  • Measuring 2 x 3 inches
  • The plastic materials are polyester and acetate in the thin set (0-10 mils) and unplasticized vinyl in the thick set (10-60 mils)

Technical Attributes

  • Data
    Each set contains six chips of the following nominal film thicknesses:
    Thin Set:0.5, 1.0, 3.0, 5.0, 7.5, 10 mils
    Thick Set:10, 20, 30, 40, 50, 60 mils / Each film is also identifiable by its distinctive color.

Procedure

Operating instructions for measuring instruments will normally include calibration procedures. In the absence of written procedures, the user is advised to contact the instrument manufacturer. In general, proximity type instruments are most reliably calibrated for a given measurement by selecting two standard films which “bracket” (one above, one below) the expected film thickness to be measured. By first causing the instrument to read both standards correctly over a bare surface (the same or similar to the coated one), then the unknown reading should be correctly made. Of course, the narrower the range between the two standards the lesser are the possibilities of errors due to instrumental non-linearities.

An optical-mechanical gauge may be calibrated by the owner following a simple verification procedure on a standard test panel. If observed thicknesses deviate from “true” values, a calibration curve can be drawn to specify necessary corrections. Restoration of “zero” correction, if desired, will usually require attention by the manufacturer’s shop to re-establish precise alignment.

Delivery Content

Each set contains six chips of the following nominal film thicknesses:

11299 - Thin Set: 0.5, 1.0, 3.0, 5.0, 7.5, 10 mils

11300 - Thick Set:10, 20, 30, 40, 50, 60 mils / Each film is also identifiable by its distinctive color.