Surface Profile and Coating Thickness Gage

Standards: ISO 2808-4B; ASTM D 4417-B; JIS K 5600-1-7; BS 3900-C5

The TQC Surface Profile & Coating Thickness Gage is a combination gage that can be equipped with two different tips, one for surface roughness and another for coating thickness.

Comes complete with two tips and a glass calibration plate, all in a leather pouch.




TECHNICAL DATA USE MORE INFO PRICE LIST
 

 

Technical Data

top
Technical Data
Range: 0-3.4mm / 0-0.13 inch
Resolution: 1μm / 0.04 mil
Accuracy: ±5μm / 0.2 mil
Thread: M2.5 x 0.45
Stem Diameter: 8mm / 0.3 inch
TipPoint: 60° angle


How to Perform Measurement

top
Measuring Roughness
  1. Press the On/Off button to switch the gage on.
  2. Check if the right tip is chosen. (the sharp needle tip is suitable for measuring roughness)
  3. Choose parameter by pressing the IN/MM button.
  4. Place the needle of the gage on the flat glass specimen (zero plate) and press the gage with the
    holder down until the base of the holder stands firmly on the zero plate.
  5. Press the ZERO button to make the instrument read zero.
  6. Place the needle gently on the blasted surface and press the base of the gage-holder firmly against the steel. DO NOT drag the instrument.
  7. Read the peak-valley value.
  8. Make 10 measurements on each desired location and determine the mean as being the profile of
    the surface.

Measuring Thickness
  1. Press the On/Off button to switch the gage on.
  2. Check if the right tip is chosen. (the round tip is suitable for measuring thickness)
  3. Choose parameter by pressing the IN/MM button.
  4. Place the needle of the gage on the flat glass specimen (zero plate) and press the gage with the holder down until the base of the holder stands firmly on the zero plate.
  5. Press the ZERO button to make the instrument read zero.
  6. Gently remove a piece of paint with a diameter of 8mm. from the surface. Try to remove the paint without damaging the underground material.
  7. Place the needle on the removed paint. Make sure the aluminum footing stands on the painted area.
  8. The Coating thickness appears on the display.

 

Maintenance

top
  • This instrument is precision-machined. Never drop it or knock it over.
  • Always clean the instrument after use.
  • Clean the instrument using a soft dry cloth. Never clean the instrument by any mechanical means such as a wire brush or abrasive paper. This may cause permanent damage, just like the use of aggressive cleaning agents can.
  • Do not use compressed air to clean the instrument.
  • Always keep the instrument in its case when not in use
  • Annual calibration recommended

 

Changing Tips

top
Changing Tips Step 1
Two tips are supplied. The sharp needle tip is suitable for measuring roughness, the round one for thickness.
Step 2
Take the tip between two fingers and turn anti-clockwise until it is loose. If the tip stays stuck, go to step 2a.
Step 2a
Only when loosening the tip anti-clockwise by hand fails, a pair of tongs may be used GENTLY. Make sure the tip remains undamaged.
Step 3
Turn the new tip clockwise until it’s stuck.
Step 4
The tips have been changed. Don’t forget to store the tip that’s not in use.

 

Prices

top
Prices subject to change    Prices listed in US dollars for domestic shipments within US, Canada, & Mexico only    Export orders require quotations - click here for a quote     For more info call us 1-800-762-2478 or 954-946-9454.
Item Number Item Prices

Instrument comes with 2 tips, a glass calibration plate & a leather pouch
SU-891023 SP1560 Surface Profile/Coating Thickness Gage $345.00
SU-891023/C SP1560 Certified Surface Profile/Coating Thickness Gage 495.00
Replacement Tips
SU-891027 Replacement tip for coating thickness 45.00
SU-891028 Replacement tip for roughness 45.00
SU-891031 Replacement Battery, LR44 Button Cell,1.5V 10.00



top

Call HAVE QUESTIONS?
    1-800-762-2478
Email PREFER TO EMAIL?
    gardner@gardco.com
  FIND PRODUCTS LEARN MORE ORDERING INFO ABOUT US
  By Category Email Registry How to Order Contact Info
  By Industry Trade Shows Request Quote History
  New Products Test Lab Info Order Form FAQ's
  Monthly Product Articles/Links Sales & Clearance Policies




  © 2018 Paul N. Gardner Company, Inc. | Contact | Policies Facebook  Twitter  YouTube  Blogger