Home  >>  Products  >>  Adhesion  >>  Hydraulic Adhesion Tester
Hate Hydraulic Adhesion Tester

The HATE Mark VII coating adhesion testers offer the coating industry new capabilities, from the lab technician and paint chemist to the field inspector and corrosion engineer. The adhesion strength specified can now be proved to have been achieved - even on and in, tubulars in the most awkward positions!



DIAGRAM TECHNICAL DATA MORE INFO PRICE LIST
 

 

Diagram

top

The design of the HATE tester is simple. Load is applied through the center of the dolly by a hydraulic piston and pin (Fig. 1). The hemispherical tip of the pin ensures a point loading of the force. Consequently, the force reacted in the dolly is a uniform, purely tensile load. The diameter of the piston bore is sized so that the area of the bore is equal to the net area of the dolly. Therefore, the pressure reacted by the dolly is the same as the pressure in the bore which is transmitted directly to a pressure gauge. This area ratio of 1:1 and a direct reading from a pressure gauge ensures repeatable accuracy. The accuracy of the HATE tester is ± 1%, this comparing to an accuracy of ± 10% for mechanical adhesion testers.

 

Techincal Data

top
Technical Data
Model Pressure Rating
PA-8735: Dual Scale 3,500 psi /250 MPa Other pressure ratings are available on request. Working range is 75% of Full Scale Deflection (F.S.D.)
Accuracy: All models have an accuracy of ± 1% of F.S.D.
Tolerances: The internal diameters of the piston bore and dolly and the dolly outside diameter are machined to a tolerance of ±1 thousandth of an inch.

 

Safely

top

Your HATE adhesion tester is intrinsically safe. By turning the handle clockwise, a finely threaded plunger decreases volume, and increases the hydraulic pressure on the oil within the body. As oil is an incompressible fluid, any leakage will cause a dramatic drop in pressure. Your HATE tester requires only 0.5 cubic centimeters to be lost for the pressure to drop from 2,000 psi to 0 psi.

With a mechanical system, as the spring stores a large amount of energy, a pull-off results in a sudden release of this energy causing a loud bang and the tester to jump. Because your HATE tester operates hydraulically, only a small amount of energy is stored, therefore a pull-off is a lot quieter.

 

Material Specs

top
Material Specs
Handle: Duroplastic
  Threaded Plunger Bright Bar Steel
Body 316 Stainless Steel
Hose Synthetic rubber tube, double wire braid reinforcement & synthetic rubber cover
Connector: Stainless steel
  Pin 4140 steel
Case Plastic
Plug Teflon
Dolly: (304 Stainless Steel)
  Outside diameter 0.75 inch
Inside diameter 0.125 inch
Adhesive: Cyanoacrylate Scotchweld 3M-20
  Curing time: 2,000 psi in 15 minutes at 25°C
3,000 psi in 30 minutes at 25°C
4,250 psi in 24 hours at 25°C
Net Weight:
3 lbs. (1.4kgs) (including case & accessories)
Gross Weight: 10lbs. (4.6kgs)
Length:
Overall length 20 inches (510 mm)

The standard length of hose is 200mm. Extra length hose can be supplied, up to a maximum of 600mm.

For testing inside pipe, a 90° elbow fitting can be specified.

Shown Right, Complete Kit with Cleaning Accessories

 

Prices

top
Prices subject to change    Prices listed in US dollars for domestic shipments within US, Canada, & Mexico only    Export orders require quotations - click here for a quote     For more info call us 1-800-762-2478 or 954-946-9454.
Item Number Item Prices

5 flat dollies, 5 Teflon plugs, 20 gram bottle of cyanoacrylate adhesive, electrical dolly cleaner, plastic case
PA-8735 Dual Scale MARK VII Adhesion Tester - Range 0-3,500 psi / 25 MPa $1,800.00
PA-8815 Flat Dolly for PA-8735 (Each) 36.00
PA-8850 Teflon Plug 10.00
PA-8860 Adhesive 28.00
PA-8870 Carrying Case (Replacement) 154.00
PA-8880 Electrical Dolly Cleaner 130.00
Concave & Convex Dollies for pipes (specify I.D. & O.D. in/mm) P.O.R.



top

Call HAVE QUESTIONS?
    1-800-762-2478
Email PREFER TO EMAIL?
    gardner@gardco.com
  FIND PRODUCTS LEARN MORE ORDERING INFO ABOUT US
  By Category Email Registry How to Order Contact Info
  By Industry Trade Shows Request Quote History
  New Products Test Lab Info Order Form FAQ's
  Monthly Product Articles/Links Sales & Clearance Policies




  © 2018 Paul N. Gardner Company, Inc. | Contact | Policies Facebook  Twitter  YouTube  Blogger