Positector 6000 FNDS Probe Dimensions

PosiTector 6000 FNDS Duplex Coatings Probe

Art. No.: 15273

Probe FNDS, Regular Cable, 0-60 mil

The PosiTector 6000 FNDS probe measures the individual thicknesses of both the paint and the zinc layers in a duplex coating system with a single reading.

Select your Model

€897.00

Enter a value between 1 and 100000000 in increments of 1.

excl. taxes & shipping cost

  • Helps in corrosion protection - Duplex coating systems use a combination of two corrosion protection systems - typically paint or powder coating over galvanized steel (hot-dip, electro or zinc spray metallizing). The resultant corrosion protection is superior to either protection system used independently.
  • FNDS probe in Duplex mode utilizes both magnetic (ferrous) and eddy current (non-ferrous) principles simultaneously to calculate and display the individual paint and zinc layer thicknesses.
    • The magnetic principle is used to measure the combined paint/zinc thickness over the ferrous substrate
    • The eddy current principle is used to measure the paint thickness over the non-ferrous zinc coating. The zinc thickness is calculated by subtracting the paint thickness from the combined paint/zinc thickness measurement.
  • When disabled, the gage will operate like a conventional combination ferrous/non-ferrous instrument (similar to PosiTector 6000 FNS probe). Ideal for measuring non-magnetic coatings over steel and non-conductive coatings over non-ferrous metal substrates.
  • Compatible with PosiTector Advanced (serial numbers 730,000+) and PosiTector Standard gage bodies with a color display. A Simple gage update may be required.

Technical Data Chart
Data
Probe ModelFNDS
Substrates:Ferrous / Non-ferrous
Range:0 - 60 mils (0 - 1500 µm)





Duplex Mode:
Minimum Zinc Thickness:0.5 mil (13 µm)

Paint Layer Accuracy:
±(0.05 mils + 1%) 0 - 2 mils
±(0.1 mils + 1%) > 2 mils

±(1 µm + 1%) 0 - 50 µm
±(2 µm + 1%) > 50 µm

Zinc Layer Accuracy:
±(0.1 mils + 1%) 0.5 - 2 mils
±(0.2 mils + 1%) > 2 mils

±(2 µm + 1%) 13 - 50 µm
±(4 µm + 1%) > 50 µm
Calibration Standards:S1 & A1
Accuracies are stated as a fixed value plus a percentage of the gage's actual reading