| Krebs Unit Viscometer Model KU-2 | ||
DIGITAL DIRECT DISPLAY OF VISCOSITY IN KREBS UNITS
Measures Paints, Coatings And Related Materials
|
|
||
| SPECIFICATIONS |
| Range | 40 to 141 KU; 32 to 1099 gm; 27 to 5,274 cP |
| Resolution | Within 0.1 KU or 1 gm |
| Accuracy | Within ±1% of full scale |
| Reproducibility | Within ±0.5% of full scale |
| Operating temperature | 10°C to 40°C |
| Spindle speed | 200 rpm ±0.1rpm |
| Measuring containers | 1 pint, 1/2 pint, 1 quart cans |
| Power line requirement | 110 or 220 volts, 50 or 60 Hz |
| References | ASTM D562, D856, D1131 |
| Shipping weight | 22 lbs. (10 kg) |
| VISCOSITY STANDARD OILS |
| VISCOSITY STANDARD | VISCOSITY CENTIPOISE | KREBS UNIT KU |
| VI-0280 | 400 | 64 |
| VI-0281 | 750 | 79 |
| VI-0282 | 940 | 84 |
| VI-0283 | 1400 | 95 |
| VI-0284 | 2000 | 106 |
| OPERATION |
Samples are tested in standard paint industry cans (1 pint, 1/2 pint, or 1 quart) held by a centering and retaining fixture. The container is filled to within 3/4" (20mm) of the top and located on the viscometer base The operating lever has three positions:. |
|||||
|
|||||
The displayed reading can be held by depressing the HOLD
switch at any time. With the viscometer in the For best reproducibility of sample measurements, the test material should be preconditioned for temperature and shear stability. |
|||||
Please note:
- The cP reading from the KU-2 will not necessarily agree with cP values
measured with other instrumentation due to:
- Unique shear rate of the KU-2 paddle spindle.
- Uniqueness of the Krebs unit scale. (The centipoise value should not be used for purposes of calibration.
- Use only Brookfield KU mineral oils for calibration check.
- The cP readings will only be valid when using the paddle spindle.
- The paste spindle may not be used to determine cP.
| available items |
| Item Number | Item |
| VI-020102 | Krebs Unit Viscometer - Standard Model, 115V |
| VI-020104 | Krebs Unit Viscometer w/ Air Purge, 115V |
| VI-020112 | Krebs Unit Viscometer - Standard Model, 230V |
| VI-020114 | Krebs Unit Viscometer with Air Purge, 230V |
| TH-02280 | Stormer ASTM 49C Thermometer (20-70°C x 0.2°) |
| TH-02280/C | Stormer ASTM 49C Thermometer w/ NIST Certificate |
| Spindles | |
| VI-0223 | Krebs Replacement Spindle |
| VI-0225 | Krebs Paste Spindle |
| Paul N. Gardner Company, Inc. • Phone 1.800.762.2478 • Fax 954.946.9309 |