| Brookfield Cap Series Cone/Plate Digital Viscometers |
CAP 1000+ AND CAP 2000+
![]() |
|
|||||||||||||||||||||||||||||||||||||||||||
| Brookfield CAP 1000+ Viscometer | ||||||||||||||||||||||||||||||||||||||||||||
| Applications |
The CAP 1000+ Viscometer is appropriate for a wide range of materials, including Paints &
Coatings, Adhesives & Sealants, Polymers & Resins, and more.
| CAP 1000+ Cone Viscosity Ranges (Poise) | |||||||||
| Spindle | Shear Rate |
Sample Volume |
Cone Angle |
Cone Radius |
Viscosity 5 RPM (Poise) |
Viscosity 50 RPM (Poise) |
Viscosity 500 RPM (Poise) |
Viscosity 750 RPM (Poise) |
Viscosity 900 RPM (Poise) |
| CAP-01 | 13.3N sec-1 | 67µl | 0.45° | 1.511 cm | 37.5 - 375 | 3.75 - 37.5 | 0.375 - 3.75 | 0.25 - 2.5 | 0.2 - 2.0 |
| CAP-02 | 13.3N sec-1 | 38 µl | 0.45° | 1.200 cm | 75 - 750 | 7.5 - 75 | 0.75 - 7.5 | 0.5 - 5 | 0.4 - 4.0 |
| CAP-03 | 13.3N sec-1 | 24 µl | 0.45° | 0.953 cm | 150 - 1500 | 15 - 150 | 1.5 - 15 | 1 - 10 | 0.8 - 8.0 |
| CAP-04 | 3.3N sec-1 | 134 µl | 1.8° | 1.200 cm | 300 - 3000 | 30 - 300 | 3 - 30 | 2 - 20 | 1 - 16 |
| CAP-05 | 3.3N sec-1 | 67 µl | 1.8° | 0.953 cm | 600 - 6000 | 60 - 600 | 6 - 60 | 4 - 40 | 3 -33 |
| CAP-06 | 3.3N sec-1 | 30 µl | 1.8° | 0.702 cm | 1500 - 15000 | 150 - 1500 | 15 - 150 | 10 - 100 | 8 - 83 |
| CAP-07 | 2.0N sec-1 | 1700 µl | 3.0° | 2.399 cm | 62.5 - 625 | 6.25 - 62.5 | 0.625 - 6.25 | 0.42 - 4.2 | 0.35 - 3.47 |
| CAP-08 | 2.0N sec-1 | 400 µl | 3.0° | 1.511 cm | 250 - 2500 | 25 - 250 | 2.5 - 25 | 1.67 - 16.7 | 1.39 - 13.9 |
| CAP-09 | 2.0N sec-1 | 100 µl | 3.0° | 0.953 cm | 1000 - 10000 | 100 - 1000 | 10 - 100 | 6.67 - 66.7 | 5.5 - 56 |
| CAP-10 | 5.0N sec-1 | 170 µl | 1.2° | 1.511 cm | 100 - 1000 | 10 - 100 | 1 - 10 | 0.67 - 6.67 | 0.56- 5.56 |
| Note: Viscosity ranges shown above are for illustration. The exact range will depend upon instrument configuration. | |||||||||
| cap 2000+ |
|
![]() |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| software |
![]() |
|
| technical specifications |
| CAP 1000+ | CAP 2000+ | ||
| Viscosity Range (Poise): (Dependent on cone selected) |
Min./Max. | See Charts | Viscosity Ranges |
| Speeds: | RPM | 900/750 | 5 - 1K |
| No. of Increments | 2 | 995 | |
| Spindles (included): | 1 | 1 | |
| Data Displayed: | Read-Out | Digital | Digital |
| Torque % | |||
| P (Poise) | |||
| mPa |
|||
| Shear Rate | |||
| Shear Stress | |||
| Temperature F° / C° | |||
| Output: | 0 - 1 Vdc Chart Recorder | ||
| Analog torque/temperature | |||
| Printer Output | |||
| RS-232 PC: Data Output | |||
| RS-232 PC: Full Computer Control | |||
| Program Features: | Time to Torque | ||
| Time to Stop | |||
| Customize 4 Programs | |||
| Customize 10 Programs | |||
| Accessories: | Water Baths | ||
| Thermosel Heaters | |||
| Helipath Stand | |||
| Spiral Adapter | |||
| Small Sample Adapter | |||
| UL Adapter | |||
| DIN Adapter | |||
| Standard Spindles | |||
| Cone Spindles | 45 | 45 | |
| T-Bar Spindles | |||
| Connections & Extensions | |||
| Viscosity Standard | 28 | 28 | |
| Software: | DV Loader | ||
| RheoLoader | |||
| Wingather | |||
| Rheocalc | |||
| Capcalc | |||
| Rheo2000 | |||
| RheoVision | |||
| cone spindle range table |
| Item No. | Cone
Spindle No. |
Cone
Angle |
Shear
Rate (sec-1) |
Full-Scale Viscosity Range (Poise) |
Sample
Size (µl) |
|
| CAP 1000+ at 750 RPM |
CAP 1000+ at 900 RPM |
|||||
| VI-0471 | 01 | 0.45° | 13.3N | .25 - 2.5 | 0.21 - 2.08 | 67 |
| VI-0472 | 02 | 0.45° | 13.3N | .5 - 5.0 | 0.42 - 4.17 | 38 |
| VI-0473 | 03 | 0.45° | 13.3N | 1.0 - 10.0 | 0.83 - 8.33 | 24 |
| VI-0474 | 04 | 1.8° | 3.3N | 2.0 - 20.0 | 1.67 - 16.7 | 134 |
| VI-0475 | 05 | 1.8° | 3.3N | 4.0 - 40.0 | 3.33 - 33.3 | 67 |
| VI-0476 | 06 | 1.8° | 3.3N | 10.0 - 100.0 | 8.33 - 83.3 | 30 |
| VI-0477 | 07 | 3.0° | 2.0N | .42 - 4.2 | .35 - 3.5 | 1700 |
| VI-0478 | 08 | 3.0° | 2.0N | 1.66 - 16.6 | 1.4 - 14 | 400 |
| VI-0479 | 09 | 3.0° | 2.0N | 6.67 - 66.7 | 5.5 - 55 | 100 |
| VI-0470 | 10 | 1.2° | 5.0N | 0.67 - 6.7 | 0.5 - 5.0 | 170 |
| N = RPM | ||||||
| cap ordering information |
The CAP Viscometer is supplied with choice of one cone spindle. Please specify line voltage and frequency when ordering. Choice of operating temperature range must also be specified with order: 5° C to 75° C, or 50° C to 200° C. Accessories include: Certified Viscosity Standard Fluids which are necessary for performing calibration tests; additional standard cone spindles or custom designed spindles; Brookfield CAPCALC application software. All Brookfield Viscometers are guaranteed for one year against defects in materials and workmanship. |
| helpful hints |
| Need Extended Range? | |
| Order additional cones or optional torque range. | |
| Order Brookfield Viscosity Standards to verify calibration | |
| Specific Standards for each cone are available. | |
| You can turn your CAP2000+ into a "Rheometer" | |
| Order CAPCALC software. | |
| |
| ITEMS AND options |
|
|
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| For Calibratimg CAP Series cones each spindle has it's own
fluid E.g. CAP1L fluid is for use with CAP 01 spindle on a low temperature CAP viscomete.r The Fluid Part number indicates the specific CAP spindle which isused with the fluid. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Home Viscosity How To Order Top Of The Page | Paul N. Gardner Company, Inc. • Phone 1.800.762.2478 • Fax 954.946.9309 |