Home > Products > pH/Conductivity > 310 Oil Content Analyzer
Oil Content Analyzer - Model OCMA-310

The new OCMA-310 Oil Content Analyzer can assess the amount of trace hydrocarbons in soil or water, as well as check for residual hydrocarbons on pre-cleaned products. The extraction of the oil can be done by Freon 113 and S-316. The S-316 solvent used by the unit may be recycled by HORIBA’s optional solvent reclaimer, making the OCMA-310 very environmentally-friendly. Furthermore, the OCCMA-
310 is quick, easy to use and accurate and reliable.


Manual Control - The two valve system lets you make sure that your sample is measured properly. Inspect your sample and press the extract button to initiate making and separation. You can visually confirm mixing and separation through the sample window. Open the upper valve (extraction valve), and the sample will flow into the measuring cell. An oil/water separation filter prevents water from entering
the cell. An overflow system discharges any excess sample. The lower valve (drainage valve0 allows you to drain the sample after it has been measured.

0-200 mg/l Dynamic Range - The OCMA-310’s pyroelectric sensor gives you a full dynamic range from 0-200 mg/L. This eliminates the need for range-switching and provides fast, accurate measurements for all types of samples, regardless of concentration levels.

Quick Start-up - The OCMA-310 warms up in less than 30 minutes, enabling you to get started quickly.

No Optical Adjustments - The troublesome optical adjustments commonly required by conventional pneumatic NDIR analyzers are no longer necessary.

Auto Hold Function - The unit’s internal computer automatically determines when the measured value has stabilized and maintains that value on the display.

Self-Diagnosis - The OCMA-310 monitors itself for problems such as motor malfunction, deterioration of the measuring cell’s light source, and electrical trouble. Should anything out of the ordinary occur, an alarm light will turn on, and an appropriate error message will appear on the operation display panel.

Oil/water Separation Filter - The OCMA-310 is equipped with an oil/water separation filter which prevents water and dirt from entering the measuring cell. This prevents the cell from being damaged, while insuring that the sample being measured is not contaminated.


Flow Pattern During Measurement
Stage 1
Check that the drainage valve is closed, then open the extraction valve. The sample will flow into the measurement cell where it is measured.
 
Stage 2
Close the extraction valve and open the drainage valve. The previously measured sample will drain out of the unit.
 
Flow pattern when draining excess sample or solvent
Open both the extraction and drainage valves simultaneously. All the fluid will drain out of the unit.
 
Complete dispersion, mixing and extraction
The extraction tank uses HORIBA’s specially designed agitator blades. Pushing the EXTRACT button initiates a powerful rotary dispersion and mixing action. To assure proper measurement of your oil sample, the extraction time can be preset in 10 second increments to any desired interval from 10 seconds to 10 minutes.

Recommended Applications
Environment: Surveying environmental water quality and hazardous waste sites
Industry: Monitoring of waste-water influent and effluent
Marine Transportation: Checking bilge and ballast discharge
Petroleum Processing: Monitoring the efficiency of oil/water separation process
Oil Depots: Monitoring the discharge produced when cleaning storage tanks
Automotive: Monitoring waste-water discharge from service stations
Quality Assurance: Measuring residual oil on textiles, metal parts, etc.

Technical Data
Application: Oil contamination of fresh and salt water; soil analysis
Principle: Solvent extraction, NDIR analysis (infrared spectroscopy)
Detector: Pyroelectric sensor
Measuring Range: 0 to 200 mg/L
Resolution: 0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg;L
Repeatability: 0 to 9.9 mg/L; ±0.2 mg/L ±1 digit
10.0 to 99.9 mg/L; ±2.0 mg/L ±1 digit
100 to 200 mg/:’ ±4 mg/L ±1 digit
Measurement: Manually controlled
Calibration: One touch calibration after the calibration standard is introduced to the instrument
Extraction Solvent: HORIBA S-316 solvent (Standard 3 accessories) *use only S-316, not other solvent
Sample/Solvent Volume: 20 ml, ratio of sample to solvent is 2:1
Extraction Method: Built-in extractor for oily water samples
Display: Measured value; 3 digits LCD with back light
Message: Character display LCD with back-light (16 x 2 characters), English or German
Functions: Data memory up to 50 Data, self diagnostics, auto hold function, interactive operation, calendar clock
Output: RS-232C, parallel printer port. Option: Analog output 0 to 200 mg/L; 0 to 1V
Ambient Temperature: 0 to 40°C
Power Requirements: 100 to 120V AC ±10%, 50/60 Hz, 80 VA; 200 to 240 V AC ±10T, 50/60 Hz, 120 VA
Dimensions: 200(W) x 315(D) x 342 (H) mm
Weight: Approx 7 kg

Available Items - Literature
Prices subject to change    Prices listed in US dollars for domestic shipments within US, Canada, & Mexico only    Export orders require quotations - click here for a quote     For more info call us 1-800-762-2478 or 954-946-9454.
Item Number Item Prices
WC-380310 OCMA-310 Oil Content Analyzer $7,500.00
WC-3805 S-316 Solvent in 1.5 kg Bottle 360.00
WC-389011 SR-305 Solvent Reclaimer. Includes: O-Rings (2/set) for Model SR-305 using S-316 Solvents, Filters (2/set), Activated Carbon (500g) and Activated Alumina (90g) 1,650.00

Compare Models OCMA-310 and OCMA-350
 
  OCMA-310 OCMA-350

Sampling Method:

1) Extraction tank automates
2) After mixing, S-316 layer passes through a water separation filter sample cell and placed in the IR path
3) Maintenance required to change filter periodically and corroded parts as necessary


1) Operator performs extraction extraction process
2) S-316 layer is poured into a quartz
3) No maintenance required

Calibration: Manual zero and span Manual zero and span
Measuring Range: 0-200mg/L dynamic range for oil in water samples only 0-20mg/L (oil in water)
0-1,000mg/kg (for Oil measurements in Soil)
0-1.0Abs (for Indicating Absorption)
Repeatability: 0-9.9mg/L: ±0.2mg/L ±1.0 digit
10-99.9mg/L: ±2.0 mg/L ±1.0 digit
100-200mg/L: ±4.0mg/L ±1.0 digit
0-9.9mg/L: ±0.4mg/L ±1.0 digit
10-99.9mg/L: ±2.0mg/L ±1.0 digit
100-200mg/L: ±4.0mg/L ±1.0 digit
0-1Abs: ±1.0%FS