| Oil Content Analyzer - Model OCMA-350 |
Push-button simplicity, perfect for field analysis
![]() |
| |
| Item Number WC-3801 |
| Recommended Applications |
| Special Features |
The OCMA-350 has various special features to assure accurate measurements.
| Stable readout | |
| The unit's data evaluation function analyzes the readings, determines when the value has stabilized, and displays that stable value. | |
| Self-monitoring | |
| Messages on the LCD screen inform you of any electrical malfunctions, irregularities in measurement, and deterioration of parts, so that you can always be assured of proper functioning. | |
| Simplified record keeping | |
| The OCMA-350 keeps a record of the time and date of measurement along with each data set it records. When printing out, it provides you with both the time and the measurement, allowing for easy record keeping. |
|
|||
| Easy-to-read backlit LCD | |||
| The backlit LCD is easy to read under any conditions, including bright sunlight. The digital display eliminates individual reading differences. | |||
| Three measurement units | |||
| The OCMA-350 measures in mg/I for oil in water, mg/kg for oil in soil, and Abs for the absorbency of oil, so you can always choose the proper unit for your sample measurement. | |||
| Cell control | ![]() |
|
![]() |
All the right connections |
|
| S-316 SOLVENT |
| ![]() |
| S-316 Solvent |
| Specifications |
| Application: | Oil contamination of fresh and salt water; soil analysis | ||
| Principle: | Solvent extraction, NDIR analysis (infrared spectroscopy) | ||
| Detector: | Pyroelectric sensor | ||
| Cell: | Quartz cell, cell length, 20 mm | ||
| Measuring range and unit: | 0 to 200 mg/l, 0 to 1000 mg/kg, 0 to 1 Abs. | ||
| Resolution: | mg/l: | 0 to 99.9 mg/l; 0.1 mg/l | |
| 100 to 200 mg/l; 1 mg/l | |||
| mg/kg: | 0 to 99.9 mg/kg; 0.01 mg/kg | ||
| 10.0 to 99.9 mg/kg; 0.1 mg/kg | |||
| 100 to 1000 mg/kg; 1 mg/kg | |||
| Abs: | 0 to 1000 Abs; 0.001 Abs | ||
| Repeatability: | mg/l: | 0 to 9.9 mg/l; ±0.4 mg/l ±1 digit | |
| 10.0 to 99.9 mg/l; ±2.0 mg/l ±1 digit | |||
| 100 to 200 mg/l; ±4 mg/l ±1 digit | |||
| Abs: | ±1% F.S. | ||
| Measurement: | Manually controlled | ||
| Calibration: | Automatic calibration after the calibration standard is introduced to the instrument | ||
| Extraction solvent: | S-316 solvent | ||
| Sample/solvent volume | Approximately 6.5 ml, ratio of sample to solvent is 1:1 | ||
| Display: | Measured value: | 3 1/2 digits LCD with backlight | |
| Message: | Character display LCD with backlight (16 x 2 characters) | ||
| Functions: | Self diagnostics, Auto hold function, Calendar clock | ||
| Output: | RS-232C and centronics printer port | ||
| Ambient temperature: | 0 to 40°C | ||
| Power requirements: | 100 to 240 V AC ±10%, 50/60 Hz, 60 VA | ||
| Dimensions: | 7.9(H) x 9.8(W) x 11 .0(D) inches (200(H) x 250(W) x 285(D) mm) | ||
| Weight: | Approximately 11 lb (5 kg) | ||
| Available Items |
| Item Number | Item |
| WC-3801 | OCMA -350 Oil Content Analyzer. Includes: B-Heavy Oil, Quartz Cell, Cell cap, 2amp Fuse, power cord, 10ml and 25μl syringes and Manual |
| WC-3805 | S-316 Solvent in 1.5 kg Bottle |
| WC-389011 | SR-305 Solvent Reclaimer Includes: O-Rings, 2/set for Model SR-305 using S-316 Filters, 2/set Activated Carbon, 500g and Activated Alumina, 90g. |
| Paul N. Gardner Company, Inc. • Phone 1.800.762.2478 • Fax 954.946.9309 |