| Pensky-Martens Closed Cup Flash Tester |
Conforms to ASTM D93, FTM 791-1102, AASHTO T73-811
|
Determines flash points of liquid products, including fuel oils, lube oils, suspensions of solids, and liquids that tend to form a surface film under test conditions. Widely used in shipping and safety regulations, detection of contamination by volatile and flammable materials in fuel oils and lubricating oils, and in characterization of hazardous waste samples. Smooth operating cover mechanism slides shutter open and applies test flame at the turn of a knob. Cover fits over brass test cup and includes pilot flame, test flame reference bead, built-in stirrer and plated brass thermometer ferrule. Electrically heated model is equipped with a 750W nickel-chromium heater with stepless variable control for accurate, repeatable temperature rate of rise settings per specifications. Heater unit is enclosed in a stainless steel housing with cooling vents and line cord receptacle and switch for optional stirrer motor. Gas heated model has a built-in nickel plated brass natural gas burner, or can be supplied with an artificial gas burner or liquid propane burner (specify when ordering). Both models are mounted on a sturdy cast iron base. |
|
|
| Stirrer Motor |
• Two speed operation Slow speed gear motor with twin pulley rotates stirrer of Pensky-Martens Closed Cup Flash Tester at 115 RPM for the basic procedure or at 250 RPM for suspensions of solids and highly viscous samples. Chuck accepts flexible stirrer shaft of flash tester. Includes adjustable support bracket and mounting rod with threaded end for easy installation in base of K16200/K16270 or K16000 flash tester, and three-conductor line cord and plug. |
| Specifications |
| Conforms to the Specifications of: | ASTM D93, FTM 791-1102, AASHTO T73-811, IP 34, DIN 51758, ISO 2719 |
| Length x Width x Height, in.(cm): | 9 1/2 x 8x 22 1/2 (24 x 20 x 57) w/optional stirrer motor installed |
| Net Weight: | K16000: 26lbs (11.8kg) |
| K16200/K16270: 34 lbs. (15.4kg) | |
| Shipping Weight: | 40 lbs. (18.1kg) *includes accessory stirrer motor |
| Includes: | Brass test cup with heat resistant handle. |
| Order stirrer motor and thermometers separately. |
| Available Items |
| Item Number | Item |
| FP-K16200 | Pensky-Martens Closed Cup Flash Tester, Electrically Heated 115V 50/60 Hz |
| FP-K16270 | Pensky-Martens Closed Cup Flash Tester Electrically Heated 220-240V, 50/60 Hz |
| FP-K16000 | Pensky-Martens Closed Cup Flash Tester, Gas Heated |
| FP-K16220 | Stirrer Motor for Pensky-Martens Closed Tester 115V 50/60 Hz |
| FP-K16229 | Stirrer Motor for Pensky-Martens Closed Tester 220-240V 50/60 Hz |
| Accessories & Replacement Parts | |
| FP-K16010 | Cover Assembly includes: shutter, flame exposure device, stirrer and thermometer ferrule |
| FP-K16020 | Brass Test Cup with heat resistant handle |
| FP-K16040 | Natural Gas Burner; nickel plated brass |
| FP-09F | ASTM 9F Thermometer Range: 20 to 230°F |
| FP-09C | ASTM 9C Thermometer Range: -5 to +110°C |
| FP-10F | ASTM 10F Thermometer Range: 200 to 700°F |
| FP-10C | ASTM 10C Thermometer Range: 90 to 370°C |
| Tag Closed Cup Flash Tester |
Conforms to ASTM D56, FTM 791-1101, and related specifications
■ Gas, electrical or alcohol heating Determines flash points of liquid products, except cutback
|
|
|
| Specifications |
| Conforms to the Specifications of: | ASTM D56, FTM 791-1101, iP 304 |
| Length x Width x Height, in.(cm)*: | 5 x 5 x 16 (13x13x41)*with thermometers inserted |
| Net Weight: | 7 lbs (3.2 kg) |
| Shipping Weight: | 8 lbs (3.6 kg) |
| Dimensions: | 12x10x11" (31x25x28 cm) |
| Includes heater unit with electric heater or gas or alcohol burner. All models require (2) two thermometers. (See Accessories List) |
| Available Items |
| Item Number | Item |
FP-K14600 |
Tag Closed Cup Flash Tester, Electrically Heated, 115V 50/60 Hz |
| FP-K14670 | Tag Closed Cup Flash Tester, Electrically Heated, 220-240V 50/60 Hz |
| FP-K14690 | Tag Closed Cup Flash Tester, Gas Heated |
| Accessories & Replacement Parts | |
| FP-K14510 | Cover Assembly, includes slide shutter, burner and thermometer ferrules |
| FP-K14520 | Brass Test Cup with heat resistant fiber handle |
| FP-09F | ASTM 9F Thermometer, Range: 20 to 230°F |
| FP-09C | ASTM 9C Thermometer, Range: -5 to +110°C |
| FP-57F | ASTM 57F Thermometer, Range: -4 to + 122°F |
| FP-57C | ASTM 57C Thermometer, Range: -20 to +50°C |
| Tag Open Cup Flash Tester |
Conforms to ASTM D1310, D3143 specifications
■ Choice of gas, electrical or alcohol heating For liquid products, including non-petroleum products, having flash and fire points of up to 325°F(163°C). Suitable for cutback asphalts having flash points of less than 200°F(93°C). Includes Pyrex™ sample cup, plated brass liquid bath with constant level overflow, pivoting ignition taper with pilot light and reference bead, pivoting thermometer holder, heater and cast aluminum base. Electrically heated model is equipped with a stepless variable transformer for accurate control of temperature rate of rise per specifications. Also available with gas or alcohol burner. |
|
|
| Specifications |
| Conforms to the Specifications of: | ASTM D1310, D3143 |
| Length x Width x Height, in.(cm)*: | 10x7x17 (25x18x43)*with thermometer inserted |
| Net Weight: | 71/2 lbs (3.4kg) |
| Shipping Weight: | 91/2lbs(4.3kg) |
| Dimensions: | 16x12x8" (41x30x20cm) |
| Includes heater unit with electric heater or gas or alcohol burner. Order thermometers, leveling device and draft shield separately. All models require (2) two thermometers. |
| Available Items |
| Item Number | Item |
| FP-K15600 | Tag Open Cup Flash Tester, Electrically Heated 115V 50/60 Hz |
| FP-K15670 | Tag Open Cup Flash Tester, Electrically Heated 220-240V 50/60 |
| FP-K15690 | Tag Open Cup Flash Tester, Gas Heated |
| Accessories & Replacement Parts | |
| FP-K15610 | Leveling Device for proper adjustment of sample level in test cup. Meets ASTM specifications. Polished aluminum |
| FP-K15620 | Draft Shield, steel with flat black interior per ASTM specifications |
| FP-K15520 | Pyrex™ Sample Cup |
| Thermometers | |
| FP-33F | ASTM 33F Thermometer Range: -36.5 to +107.5°F |
| FP-33C | ASTM 33C Thermometer Range: -38 to +42°C |
| FP-09F | ASTM 9F Thermometer Range: 20 to 230°F |
| FP-09C | ASTM 9C Thermometer Range: -5 to 110°C |
| FP-35F | ASTM 35F Thermometer Range: 194 to 338°F |
| FP-35C | ASTM 35C Thermometer Range: 90 to 170°C |
| Cleveland Open Cup Flash Tester |
Conforms to ASTM D92, FTM 791-1103, AASHTO T48 and related specifications
■ Choice of electric or gas heater Determines flash and fire points of all petroleum products except fuel oils and those having flash points below 175°F(79°C). Consists of test flame applicator, brass test cup, thermometer support, heating plate and gas or electric heater. Applicator is precisely aligned per specifications, and pivots for convenient test flame application at specified temperature intervals. Hinged thermometer support raises to facilitate placement and removal of test cup. Adjust flame size using built-in needle valve and comparison bead. Electrically heated model is equipped with a 750W nickel-chromium heater with a variable stepless transformer for accurate control of temperature rate of rise per specifications. Gas heated model has a built-in natural gas burner, or can be supplied with an artificial gas burner or liquid propane gas burner—specify when ordering. Mounted on a sturdy cast iron base. Heater unit of electrically heated model is enclosed in a stainless steel housing with cooling vents. Test flame applicator and thermometer support constructed of machined nickel plated brass. |
|
|
| Specifications |
| Conforms to the Specifications of: | ASTM D92, FTM 791-1103, FTM 141-4294, AASHTO T48, ANSZ-11.6, IP 36, DIN 51376, ISO 2592 |
| Length x Width x Height, in.(cm)*: | K13900/K13990: 10x5 1/2x14(25x14x36) |
| Net Weight: | K13900/K13990:17 1/2 lbs (7.9kg) |
| Shipping Weight: | 21 lbs (9.5kg) |
| Includes brass test cup with heat resistant handle, and gas inlet tube. All models require (1) one thermometer. Order thermometer separately. |
| Available Items |
| Item Number | Item |
| FP-K13900 | Cleveland Open Flash Tester, Electrically Heated 115V 50/60 Hz |
| FP-K13990 | Cleveland Open Flash Tester, Electrically Heated 220-240V 50/60 Hz |
| FP-K14000 | Cleveland Open Flash Cup / wood handle (precision machined brass with heat resistant handle) |
| FP-11F | ASTM 11F Thermometer Range: 20 to 760°F |
| FP-11C | ASTM 11C Thermometer Range: -6 to 400°C |