Site Map    Contact Us
Search Our Site

 

 

 

Pensky-Martens Closed Cup Flash Tester

Conforms to ASTM D93, FTM 791-1102, AASHTO T73-811

Determines flash points of liquid products, including fuel oils, lube oils, suspensions of solids, and liquids that tend to form a surface film under test conditions. Widely used in shipping and safety regulations, detection of contamination by volatile and flammable materials in fuel oils and lubricating oils, and in characterization of hazardous waste samples.

Smooth operating cover mechanism slides shutter open and applies test flame at the turn of a knob. Cover fits over brass test cup and includes pilot flame, test flame reference bead, built-in stirrer and plated brass thermometer ferrule.

Electrically heated model is equipped with a 750W nickel-chromium heater with stepless variable control for accurate, repeatable temperature rate of rise settings per specifications. Heater unit is enclosed in a stainless steel housing with cooling vents and line cord receptacle and switch for optional stirrer motor. Gas heated model has a built-in nickel plated brass natural gas burner, or can be supplied with an artificial gas burner or liquid propane burner (specify when ordering). Both models are mounted on a sturdy cast iron base.

Pensky-Martens Closed Cup Flash Tester

 

 

Stirrer Motor

 

• Two speed operation
• Mounts in base of Pensky-Martens tester

Slow speed gear motor with twin pulley rotates stirrer of Pensky-Martens Closed Cup Flash Tester at 115 RPM for the basic procedure or at 250 RPM for suspensions of solids and highly viscous samples. Chuck accepts flexible stirrer shaft of flash tester. Includes adjustable support bracket and mounting rod with threaded end for easy installation in base of K16200/K16270 or K16000 flash tester, and three-conductor line cord and plug.

 

 

Specifications


Conforms to the Specifications of: ASTM D93, FTM 791-1102, AASHTO T73-811, IP 34, DIN 51758, ISO 2719
Length x Width x Height, in.(cm):

9 1/2 x 8x 22 1/2 (24 x 20 x 57) w/optional stirrer motor installed

Net Weight: K16000: 26lbs (11.8kg)
K16200/K16270: 34 lbs. (15.4kg)
Shipping Weight: 40 lbs. (18.1kg) *includes accessory stirrer motor
Includes: Brass test cup with heat resistant handle.

 

 

Order stirrer motor and thermometers separately.

 

Available Items
Item Number Item
FP-K16200 Pensky-Martens Closed Cup Flash Tester, Electrically Heated 115V 50/60 Hz
FP-K16270 Pensky-Martens Closed Cup Flash Tester Electrically Heated 220-240V, 50/60 Hz
FP-K16000 Pensky-Martens Closed Cup Flash Tester, Gas Heated
FP-K16220 Stirrer Motor for Pensky-Martens Closed Tester 115V 50/60 Hz
FP-K16229 Stirrer Motor for Pensky-Martens Closed Tester 220-240V 50/60 Hz
Accessories & Replacement Parts
FP-K16010 Cover Assembly includes: shutter, flame exposure device, stirrer and thermometer ferrule
FP-K16020 Brass Test Cup with heat resistant handle
FP-K16040 Natural Gas Burner; nickel plated brass
FP-09F ASTM 9F Thermometer Range: 20 to 230°F
FP-09C ASTM 9C Thermometer Range: -5 to +110°C
FP-10F ASTM 10F Thermometer Range: 200 to 700°F
FP-10C ASTM 10C Thermometer Range: 90 to 370°C

 

 

Return to the Top of the Page...

 

 

Tag Closed Cup Flash Tester

Conforms to ASTM D56, FTM 791-1101, and related specifications

■ Gas, electrical or alcohol heating
■ Flash points below 200°F (93°C)

Determines flash points of liquid products, except cutback
asphalts, having flash points below 200°F (93°C). (For viscous products, products tending to form a surface film during testing and materials which contain suspended solids, use the Pensky Martens flash tester. For cutback asphalts use the Tag Open Cup tester).


Features stepless variable heat control with reference dial for convenient, accurate repeat setting of temperature rate of rise per specifications. Also available with gas or alcohol burner instead of electric heater. Precision machined cover mechanism simultaneously opens slide shutter and applies test flame to sample at the turn of a knob. Includes liquid bath with constant level overflow, brass test cup, plated brass thermometer ferrules and test flame reference bead. Bath and cover mechanism are constructed of plated brass. Heater is enclosed in a cast aluminum base assembly.

Tag Closed Cup Flash Tester

 

 

Specifications


Conforms to the Specifications of: ASTM D56, FTM 791-1101, iP 304
Length x Width x Height, in.(cm)*:

5 x 5 x 16 (13x13x41)*with thermometers inserted

Net Weight: 7 lbs (3.2 kg)
Shipping Weight: 8 lbs (3.6 kg)
Dimensions: 12x10x11" (31x25x28 cm)

 

Includes heater unit with electric heater or gas or alcohol burner. All models require (2) two thermometers. (See Accessories List)

 

Available Items
Item Number Item

FP-K14600

Tag Closed Cup Flash Tester, Electrically Heated, 115V 50/60 Hz
FP-K14670 Tag Closed Cup Flash Tester, Electrically Heated, 220-240V 50/60 Hz
FP-K14690 Tag Closed Cup Flash Tester, Gas Heated
Accessories & Replacement Parts
FP-K14510 Cover Assembly, includes slide shutter, burner and thermometer ferrules
FP-K14520 Brass Test Cup with heat resistant fiber handle
FP-09F ASTM 9F Thermometer, Range: 20 to 230°F
FP-09C ASTM 9C Thermometer, Range: -5 to +110°C
FP-57F ASTM 57F Thermometer, Range: -4 to + 122°F
FP-57C ASTM 57C Thermometer, Range: -20 to +50°C

 

 

Return to the Top of the Page...

 

 

Tag Open Cup Flash Tester

Conforms to ASTM D1310, D3143 specifications

 

■ Choice of gas, electrical or alcohol heating
■ Flash and fire points up to 325°F (163°C)

For liquid products, including non-petroleum products, having flash and fire points of up to 325°F(163°C). Suitable for cutback asphalts having flash points of less than 200°F(93°C). Includes Pyrex™ sample cup, plated brass liquid bath with constant level overflow, pivoting ignition taper with pilot light and reference bead, pivoting thermometer holder, heater and cast aluminum base.

Electrically heated model is equipped with a stepless variable transformer for accurate control of temperature rate of rise per specifications. Also available with gas or alcohol burner.

Tag Open Cup Flash Tester

 

 

Specifications


Conforms to the Specifications of: ASTM D1310, D3143
Length x Width x Height, in.(cm)*:

10x7x17 (25x18x43)*with thermometer inserted

Net Weight: 71/2 lbs (3.4kg)
Shipping Weight: 91/2lbs(4.3kg)
Dimensions: 16x12x8" (41x30x20cm)

 

Includes heater unit with electric heater or gas or alcohol burner. Order thermometers, leveling device and draft shield separately. All models require (2) two thermometers.

 

Available Items
Item Number Item
FP-K15600 Tag Open Cup Flash Tester, Electrically Heated 115V 50/60 Hz
FP-K15670 Tag Open Cup Flash Tester, Electrically Heated 220-240V 50/60
FP-K15690 Tag Open Cup Flash Tester, Gas Heated
Accessories & Replacement Parts
FP-K15610 Leveling Device for proper adjustment of sample level in test cup. Meets ASTM specifications. Polished aluminum
FP-K15620 Draft Shield, steel with flat black interior per ASTM specifications
FP-K15520 Pyrex™ Sample Cup
Thermometers
FP-33F ASTM 33F Thermometer Range: -36.5 to +107.5°F
FP-33C ASTM 33C Thermometer Range: -38 to +42°C
FP-09F ASTM 9F Thermometer Range: 20 to 230°F
FP-09C ASTM 9C Thermometer Range: -5 to 110°C
FP-35F ASTM 35F Thermometer Range: 194 to 338°F
FP-35C ASTM 35C Thermometer Range: 90 to 170°C


 

Return to the Top of the Page...

 

Cleveland Open Cup Flash Tester

Conforms to ASTM D92, FTM 791-1103, AASHTO T48 and related specifications

 

■ Choice of electric or gas heater
■ For flash points above 175°F (79°C)

Determines flash and fire points of all petroleum products except fuel oils and those having flash points below 175°F(79°C). Consists of test flame applicator, brass test cup, thermometer support, heating plate and gas or electric heater. Applicator is precisely aligned per specifications, and pivots for convenient test flame application at specified temperature intervals. Hinged thermometer support raises to facilitate placement and removal of test cup. Adjust flame size using built-in needle valve and comparison bead.

Electrically heated model is equipped with a 750W nickel-chromium heater with a variable stepless transformer for accurate control of temperature rate of rise per specifications. Gas heated model has a built-in natural gas burner, or can be supplied with an artificial gas burner or liquid propane gas burner—specify when ordering.

Mounted on a sturdy cast iron base. Heater unit of electrically heated model is enclosed in a stainless steel housing with cooling vents. Test flame applicator and thermometer support constructed of machined nickel plated brass.

Cleveland Open Cup Flash Tester

 

 

Specifications


Conforms to the Specifications of: ASTM D92, FTM 791-1103, FTM 141-4294, AASHTO T48, ANSZ-11.6, IP 36, DIN 51376, ISO 2592
Length x Width x Height, in.(cm)*:

K13900/K13990: 10x5 1/2x14(25x14x36)

Net Weight: K13900/K13990:17 1/2 lbs (7.9kg)
Shipping Weight: 21 lbs (9.5kg)

 

Includes brass test cup with heat resistant handle, and gas inlet tube. All models require (1) one thermometer. Order thermometer separately.

 

Available Items
Item Number Item
FP-K13900 Cleveland Open Flash Tester, Electrically Heated 115V 50/60 Hz
FP-K13990 Cleveland Open Flash Tester, Electrically Heated 220-240V 50/60 Hz
FP-K14000

Cleveland Open Flash Cup / wood handle (precision machined brass with heat resistant handle)

FP-11F ASTM 11F Thermometer Range: 20 to 760°F
FP-11C ASTM 11C Thermometer Range: -6 to 400°C

 

 

Home   Lab Apparatus    How To Order   |  Top Of The Page Paul N. Gardner Company, Inc. • Phone 1.800.762.2478 • Fax 954.946.9309