| SPECIFICATIONS |
| Max. Recording Time: | 9999 min., 59 secs. (167 hrs.) |
| Mm. Recording Time: | 1 second |
| Stir Speed: | 20 rpm |
| Accuracy: | ±2% |
| Electrical Requirements: | (Specify) 11Ov/60Hz or 220v/ 50Hz |
| Dimensions: | 7 1/2"L x 9"H x 5 1/2"W |
| Weight: | 4 lbs. |
| Electrical Connection: | 3-Wire Grounded Cord Set |
| PRINCIPLE OF OPERATION |
A
low torque synchronous motor rotates a specially shaped stirrer in the catalyzed
resin or elastomer. As gelation occurs, drag exceeds torque and the motor stalls.
Gel time can then be read on the LCD display.
| TO OPERATE |
Place the material sample in the unit (in paper throw-away cup), insert wire stirrer, set display to zero, and switch on. When gel occurs, read the time and discard the cup and stirrer. The motor can remain stalled overnight without harm.
| AVAILABLE ITEMS | |
| Item Number | Item |
| GT-S | Gardco Standard Gel Timer (110V) |
| GT-S-220 | Gardco Standard Gel Timer (220 V) |
| GT-WS | Wire Stirrers (100) |
| GT-CS | Paper Cups for Gel Timer (100) |