Site Map    Contact Us
Search Our Site


 


Quintsonic

Ultrasonic Coating Thickness Gage

 

Qunitsonic The new portable gauge was especially designed for non-destructive coating thickness measurement of paint, varnish, plastics and other insulation coatings applied on wood, plastics, glass, ceramics etc. as well as for polymer layers on metals. As a special feature, the gauge offers the possibility to measure the total thickness as well as the individual layers of a multi-layer system in only one measuring process. The robust gauge is appropriate for use in the laboratory, in production and on site.

 

Measuring Procedure

When the probe is placed on the coated surface, it sends sound pulses which penetrate the coating through to the base material. Provided the coatings and the substrate possess different acoustic properties, these pulses are reflected by the different surfaces and transmitted to the transducer in the probe. The various time intervals are measured and calculated by a microprocessor to give individual and total coating thickness. The time to obtain a reading is approx. 2 seconds or less. If the portable MiniPrint printer is connected all readings and statistical data can be printed out in individual memories and batches.

 

Technical Data

Application Polymer single and multi layers, etc.
Wall thickness measurement
metals:0,2…8mm
plastics: 0,2…3mm
Measuring range Single layer: 10 μm…500μm
Multi layer: total coating thickness max. 500μm
Resolution: 1 μm
Measuring uncertainty<100 μm ± (2μm + 3%*)
Measuring uncertainty >100 μm ± (2μm + 2%*) (* of reading)
Memory capacity max. 10,000 measuring values in max 500 batches
Statistical evaluation n, x, s, kvar, max, min, with time and date of print-out and reading
Limit setting with optical and acoustic warning when limits are exceeded
Interface for printer MiniPrint and RS232C
Power supply rechargeable batteries: 2 x AA NIMH (approx. 25000 measurements) and charger 90V-to 264V - (charging time: 4 hours); or 2 x 1.5V Mignon batteries (LR6)
Dimensions / Weight Gauge: 150mm x 82mm x 35mm / 150g., Probe 30mm x 45mm dia.
Ambient temperature -15°C…+55°C

 

Available items
Item Number Item
DF-7101 QuintSonic Ultrasonic Coating Thickness Gauge, w/Probe Carrying Case, QSOFT Software w/Cable, Rechargeable Batteries,Calibration Std., Coupling, Carrying Case
DF-7112 MiniPrint 4100 Printer
DF-7113 Rubber Protective Case

 

Home   Dry Film   How To Order   Top Of The Page Paul N. Gardner Company, Inc. • Phone 1.800.762.2478 • Fax 954.946.9309