Center timer on test panel making sure the arm is counter-rotated eliminating any play in the clockwise motion of the arm.
Make sure the Teflon® ball stylus impinges the painted surface. This is done by adjusting the counter weight at rear of stylus arm to a balanced position.
The 12 gram weight is then mounted on the stylus spindle extending above the rotating arm to give a calibrated weight load.
Weight load on the ball can be adjusted by operator so that results will closely coincide with the finger touch method of determining drying time. |