Home Site Map Contact Us Search Our Site
Home > Products > Drying Time/Gelation > Electronic Multicycle DTR
Gardco Electronic Multicycle

Circular Drying Time Recorder 5000 Series

Conforms To ASTM D 5895

Electronic Multicycle DTR
  • Programmable timing cycle within range from 1 minute to 96 hours
  • Microprocessor controlled crystal timing regulation
  • Vastly improved tracking definition
    Status and Time event count-up displayed on a 16 character LCD
  • Ambient temperature sensor switchable °F or °C with min., max. & current temperature displayed
  • Auto stop upon completion of time cycle
    Power failure indicator
  • Choice of 110 volt 60Hz or 220 volt 50Hz

The unit has been designed to control a motor so that its shaft driven 3/8" teflon stylus will complete one rotation in a period which can be programmed from 1 minute to 96 hours.

The temperature can be displayed in Centigrade or Fahrenheit.


Control Panel
Control Panel

On/Off Button
This button operates from any mode. When pressed, the unit will display the Temperature or, "Press Start" and the Green L.E.D. power indicator will light. Pressing the button again will turn the unit off and interrupt any test in progress, including any changes being made to the configuration data.

Menu Button
Pressing this button will take you into the configuration menus. This can only be done with the unit on and no test running. The new configuration will be stored in memory when this button is pressed so that the display returns to the start test screen.

Scroll Button
This button only works in the configuration mode. Pressing this button will advance through the hour range.

Start Button
This starts the test. The motor will continue to run and the elapsed time will be displayed until the stop or on/off button is pressed. The Red L.E.D. Motor Run indicator will also light.

Stop Button
This stops the test. The motor will stop and the L.E.D. will go out. The elapsed time of the test will be reset.


Operating Instructions
When the Unit is first powered up, the display will show "Drying Time Recorder" for 5 seconds and then show "unit Off Press Start"
Template

Center timer on test panel making sure the arm is counter-rotated eliminating any play in the clockwise motion of the arm.

Make sure the Teflon® ball stylus impinges the painted surface. This is done by adjusting the counter weight at rear of stylus arm to a balanced position.

The 12 gram weight is then mounted on the stylus spindle extending above the rotating arm to give a calibrated weight load.

Weight load on the ball can be adjusted by operator so that results will closely coincide with the finger touch method of determining drying time.


Program Cycle
Program the desired cycle in minutes via the scroll menu. After drying stage is completed, place the clear, plastic template provided over the circular area scribed by stylus and note appropriate scale number degree required for the film to cure hard dry. Calculate the drying time using the following formula:


Formula
DT = (Tc/360) (D)
DT = Drying Time in minutes
Tc = Time Cycle in minutes
D

= Degree on template


The character of the marks by the ball stylus relate to the drying of the film. Record the results and if desired, record the temperature and the wet film thickness as these factors are important in considering the drying time of films.

Technical Data
Ambient Operating Conditions: 32°F to 120°F (0-50°C), 5 to 95% humidity (No condensing)
Warranty: 1 Year
Shipping Weight: 7 lbs

Available Items - Literature
Prices subject to change    Prices listed in US dollars for domestic shipments within US, Canada, & Mexico only    Export orders require quotations - click here for a quote     For more info call us 1-800-762-2478 or 954-946-9454.
Item Number Item Prices
Trade-in for any brand Drying Time Recorder 20% for the Multi-Cycle
DT-5020 Electronic DTR $995.00
DT-5029 Template Replacement 9.00
Home    DryingTime/Gelation    How To Order    Top Of Page Paul N. Gardner Company, Inc. • 1.800.762.2478 • Fax 954.946.9309